Candela Gentle Max / Lase PRO Fault 13.2
- MedicaParts

- Feb 9
- 1 min read

Candela Gentle Fault 13.2
High Energy fault
Alex or YAG head energy of last treatment pulse is > 14% high.
Re-calibrate the laser.
Verify beam splitter, HDD.
Verify Calport readings.
If you still facing Fault " Candela Gentle Fault 13.2 " contact us
Comments