top of page

Candela Gentle Max / Lase PRO Fault 1.1

Updated: Feb 10


Handpiece Cover Pro Series


Handpiece Bubble Circuit

Candela Gentle Max / Lase PRO Fault 1.1 , The handpiece bubble circuit test failed to detect a signal change. Replace the cryogen and perform a purge. If the issue continues contact MedicaParts


Quick Tips

  • Replace cryogen canister and perform a purg

  • Press and hold on release button to release bubbles from circuit

  • Check Data Cable May be need to replace

  • Check Candela HandPiece Bubble Sensor PCB Board May need to replace




Candela HandPiece Bubble Sensor PCB Board
$160.00
Buy Now

Comments


bottom of page